Isocaproic acid, 2-oxo- (2TMS)

 

Replica Mass Spectra of Isocaproic acid, 2-oxo- (2TMS)

rotate: click+drag; translate: alt+click+drag; zoom: mousewheel; save: mol

Spectrum Details

analyteIsocaproic acid, 2-oxo- (2TMS)
analyte InChIInChI=1S/C12H26O3Si2/c1-10(2)9-11(14-16(3,4)5)12(13)15-17(6,7)8/h9-10H,1-8H3/b11-9-
analyte mass274.5
chromatogram6058if_63
citation 
authorsBoelling C, Liebig F, Erban A, Kopka J, Max Planck Institute of Molecular Plant Physiology, Department of Molecular Plant Physiology (Prof. Willmitzer L), Am Muehlenberg 1, D-14476 Golm, Germany
lib entry date19 July 2007
metabolite role 
retention time (sec)582.19
retention index (VAR5 method, n-alkanes C10–C36)1,236.8 (observed)
base peak (m/z)73
maximal intensity117,187
mass-intensity-peaks cardinality421 intensities
minimal m/z45
maximal m/z997
download JCAMP DXSpectrumJcampDx.ashx?id=46c0f812-c0d8-4854-9925-08e453780e26
download MSPSpectrumMsp.ashx?id=46c0f812-c0d8-4854-9925-08e453780e26

GC-Method

nameVAR5
citationKopka J, Max Planck Institute of Molecular Plant Physiology, Department of Molecular Plant Physiology (Prof. Willmitzer L), Am Muehlenberg 1, D-14476 Golm, Germany
attributetextdetails
deconvolutionChromaTOFOFFSET: "just above noise"; SMOOTHING: 20; PEAK WIDTH: 6; S/N: 2
derivatizationMEOX; TMSMETHOXYAMINATION: 90min at 30°C; TRIMETHYLSILYLATION: 30min at 37°C
detectorMS-TOFm/z = 70-600; scans:20/s; Pegasus III TOF mass spectrometer
extractionwater:chloroform (2:1; v/v) 15min at 70°C
gas chromatographAgilent GC 6890 
ion sourceEI70eV
mass spectrometry instrumentPegasus III TOF mass spectrometer 
metabolic inactivationshock-frozen in liquid nitrogen 
RIALKANEC10, C12, C15, C18, C19, C22, C28, C32, C36
separationGCCOLUMN:5%phenyl-95%dimethylpolysiloxane, 30+10m, ID:0.25mm, DF:0.25µm, 5PDM VF-5ms (Varian, Darmstadt, Germany); PROGRAM:iso1min 70°C, ramp 9°C/min, iso 5min 350°C; FLOW:Helium, 0.6mL/min; INJECTION:1µL, splitless, 230°C; TRANSFER:250°C; IONSOURCE:250°C
protocolpropertyvalue
extractionwater chloroform(2:1; v/v) 15min at 70°C
derivatizationderivatisationMETHOXYAMINATION, 90min at 30°C
derivatizationderivatisationTRIMETHYLSILYLATION, 30min at 37°C
separationIon source temperature250°C
separationColumn length30+10m (analytical column + guard column)
separationColumn inner diameter0.25mm
separationTemperature programisotherm 1min 70°C, ramp 9°C/min, isotherm 5min 350°C
ion sourceelectron ionization70eV
detectorLECO instrument modelPegasus III TOF mass spectrometer
detectorhighest observed m/zm/z = 600
detectorlowest observed m/zm/z = 70
detectorsampling frequency20 scans/s
detectormass analyzer typeMS-TOF
deconvolutionmoving average smoothing20
deconvolutionsoftwareChromaTOF
deconvolutionsignal-to-noise ratio2
RIRetention indexdecane
RIRetention indexdodecane
RIRetention indexpentadecane
RIRetention indexoctadecane
RIRetention indexnonadecane
RIRetention indexdocosane
RIRetention indexoctacosane
RIRetention indexdotriacontane
RIRetention indexhexatriacontane
ChromatographyChromatographyAgilent GC 6890