GOLM METABOLOME DATABASE
How would you rate the quality of this mass spectrum?
poor, not helpful 
 outstanding, most helpful.
This spectrum was 0 times rated; on average=0.0

Jasmonic acid (1MEOX) (1TMS)

 

Replica Mass Spectra of Jasmonic acid (1MEOX) (1TMS)

replicalib entry datedetectionmethodspecies
108 December 2005 MS-TOFM[MOR] 
1 spectrum(a)
rotate: click+drag; translate: alt+click+drag; zoom: mousewheel; save: mol

Spectrum Details

analyteJasmonic acid (1MEOX) (1TMS)
analyte InChI
analyte mass311.49
chromatogram4173GX63
citation 
authorsLiebig F, Erban A, Max Planck Institute of Molecular Plant Physiology, Department of Molecular Plant Physiology (Prof. Willmitzer L), Am Muehlenberg 1, D-14476 Golm, Germany
lib entry date01 February 2005
metabolite roleMETB
retention time (sec)1,134.49
retention index (VAR5 method, n-alkanes C10–C36)1,837.26 (predicted, according to Strehmel, N. et.al)
base peak (m/z)73
maximal intensity999
mass-intensity-peaks cardinality177 intensities
minimal m/z70
maximal m/z312
download JCAMP DXSpectrumJcampDx.ashx?id=b578b61e-d631-4a58-ac80-fd02e2baf7ce
download MSPSpectrumMsp.ashx?id=b578b61e-d631-4a58-ac80-fd02e2baf7ce

GC-Method

nameM[6]
citationErban A, Schauer N, Fernie AR, Kopka J (2006) In: Weckwerth W (ed): Methods in Molecular Biology Vol., Humana Press Inc., Totowa, USA
attributetextdetails
deconvolutionChromaTOF V1.00 (Pegasus Driver 1.61), LECO, St. Joseph, MI, USAOFFSET:1_SMOOTHING:3_WIDTH:6_SN:>2
derivatizationMEOX; TMSMethoxyamination:90min40°C; Trimethylsilylation:45min40°C
detectorMS-TOFMZ:70-600; SCANS:20/s; Pegasus III TOF-MS; LECO, St. Joseph, MI, USA
extractionmethanol:water (4:1; v/v); 15min70°C 
ion sourceEI70eV
metabolic inactivationcut; liquid nitrogen; -80°C 
partitioningmethanol fraction of a methanol:water:chloroform (approx. 2:2:1, v/v/v) partitioning 
RIALKANEC12, C15, C18, C19, C22, C28, C32, C36
separationGCCOLUMN:5%diphenyl-95%dimethylpolysiloxane (arylene), 30+10m, ID:0.25mm, DF:0.25microm, Rtx-5SilMS Integra-Guard, #525560 (Restek, Bad Homburg, Germany); PROGRAM:iso1min70°C, ramp9°C/min, iso5min350°C; FLOW:Helium, 0.6ml/min; INJECTION:1microL, splitless, 230°C; TRANSFER:250°C; IONSOURCE:250°C

comments on this mass spectrum

Post a Comment
Name:
 
Website:
Email:
Comments:
service last updated 31/08/2021 © 2008-2014 Golm Metabolome Database - All rights reserved
Top